HALO BioClass Columns for Peptide and Protein Separations

HALO BioClass columns are currently comprised of the HALO Peptide and HALO Protein phases for reversed-phase separations of peptides, proteins, and antibodies.   These columns are made using superficially porous, Fused-Core® particles, which were first introduced, for small molecule separations, in 2006.  The advantages of superficially porous particles for improved speed, efficiency and peak capacity have been well demonstrated and documented in the literature over the last 7+ years.  The benefits of Fused-Core particles are also pronounced for RPLC separations of larger molecules such as peptides, polypeptides, and proteins, which have much smaller diffusion coefficients and have much slower mass transfer than low MW analytes.  A larger average pore size and a thinner porous layer are also important for improved peak shape, efficiency, and recovery of large biomolecules.

In the last two years, the HALO BioClass portfolio has been expanded with the introduction of the new 400-Ångstrom pore size HALO Protein C4 and ES-C18 phases.   HALO Protein 400 Å columns employ a new particle size and geometry, using a 3.4 µm particle with a very thin 0.2 µm porous shell. The combination of the wide pore diameter and thin shell allow very fast separations of a wide range of polypeptides and proteins up to 500 kDa in molecular weight.   In Table 1 the characteristics and properties of the current portfolio of columns in the HALO BioClass family are shown.

Table 1  Properties of HALO BioClass Columns for Bioseparations

CharacteristicPeptide ES-C18Peptide ES-CNProtein C4Protein ES-C18
Pore Size (Å)160160400400
Particle Size(s)
m)
2.7 and 52.7 and 53.43.4
Surface Area (m2/g)90 and 6090 and 601515
Low pH Limit1122
High pH Limit8899
Maximum Temperature (°C)90909090
EndcappedNoYesYesYes
Ideal MW Range
Operable MW Range
MW ≤ 10kDa
MW
 15-20 kDa
MW ≤ 10kDa
MW
 ≤ 15-20 kDa
2 kDa < MW < 500 kDa2 kDa < MW < 500 kDa
Target AnalytesPeptides and
 polypeptides
Peptides and polypeptidesMonoclonal antibodies,
polypeptides and proteins

Polypeptides and
 
proteins

 

Elevated column temperature decreases mobile phase viscosity and improves mass transfer for these larger molecules.  It also improves recovery and reduces sample carryover and memory effects—particularly for hydrophobic biomolecules (e.g., amyloid peptides, membrane proteins).  The resistance to hydrolysis of the HALO Peptide and HALO Protein phases at low pH with high temperatures is derived from sterically protected bonding technology (Peptide ES-C18 and ES-CN; Protein ES-C18) or from very dense bonding and exhaustive endcapping (Protein C4).  This enhanced stability and excellent column lifetime translates to stable and reproducible retention times and selectivity. This stability permits reproducible qualitative and quantitative results, which are critically important for peptide mapping, biomarker quantitation, and intact protein analysis.  Some example separations using the HALO Peptide and Protein columns at elevated temperatures are shown below.

Fast Peptide and Protein Separations

In this first example (Figure 1) a very fast separation of various peptides and small proteins ranging in molecular weight from 238 to 13,700 Da is presented.  A complete separation of these 11 analytes is performed in less than one minute at 60 °C using a 50-mm column on a conventional HPLC system (delay volume > 1.0 mL) at a maximum pressure of 330 bar.  A 2.1 x 50 mm, 1.7 µm column run at an equivalent 1.0 mL/min flow rate at 60 °C would generate a maximum pressure of  ~650 bar (9444 psi).

Figure 1

This second example demonstrates the efficient separation of a mixture of polypeptides and proteins ranging in molecular weight from ~12 to 250 kDa in less than 10 minutes (Figure 2, black trace).  The second chromatogram (red trace) from a subsequent injection of the same sample, after over 14,500 column volumes of mobile phase, shows very little change in retention at 90 °C.  The applicability, performance and stability of the HALO Protein C4 phase are clearly demonstrated under these conditions.

Figure 2

Summary

HALO BioClass columns will deliver speed, high peak capacities, and ruggedness to your HPLC and UHPLC separations of peptides, polypeptides and proteins.  MAC-MOD Analytical is currently offering an excellent opportunity for you to try these columns.  Please refer to our web site for more information about the HALO BioClass columns and about our current offer.

HPLC LabNotes
are offered as a courtesy of MAC-MOD Analytical. If this email has been forwarded to you by one of your colleagues and you wish to subscribe to receive future HPLC LabNotes directly, please visit www.mac-mod.com and click on HPLC LabNotes on the sidebar. You may also email us at hplc@mac-mod.com and provide us with your email and street addresses.

If you prefer not to receive HPLC LabNotes, please unsubscribe by emailing hplc@mac-mod.com


Improve Resolution and Recovery of Peptides by Increasing Column Temperature
07.21.2016
How to Improve Resolution and Recovery of Peptides by Increasing Column Temperature more...
Reduce Baseline Noise For Peptide Gradient LC-UV Analysis
03.10.2016
Reduce Baseline Noise For Peptide Gradient LC-UV Analysis more...
The Moderate Polarity Power of 3: ACE Excel SuperC18
10.06.2014
This lab note describes the features and benefits of ACE SuperC18 and links to a video. more...
Advancing Selectivity With the Polar Power of 3
08.28.2014
An Application of different stationary phases having different modes of retention more...
Specific Applications for HALO Peptide ES-C18 and ES-CN Columns
07.14.2014
Some specific applications of HALO Peptide columns are shown, and advantages vs. other commercial columns given. more...
An Overview of Systematic RPLC Method Development Using Multiple Organic Modifiers and Different pH Buffers
04.05.2014
An efficient and effective strategy for reversed-phase method development for HPLC and UHPLC usually utilizes several stationary phases with different selectivities. more...
Go Ballistic...with HALO-5 Fused-Core 5-mm HPLC Columns!
05.23.2013
With some instruments and autosamplers you can use an injector program to delay the injection a predetermined amount of time. more...
HPLC and UHPLC Separations at Mid to High pH: Part 1
05.22.2013
In this LabNote we will review the findings that were published by Kirkland, Claessens and co-workers in the 1990s. more...
HPLC and UHPLC Separations at Mid to High pH: Part 2
05.22.2013
This month's LabNote continues our discussion from January of the use of higher pH for method development and analysis. more...